M-21 Variable area flowmeter Series Instruction Manual C-MI-M21 Rev.: 2 English version The following instruction manuals are enclosed: AMD Lim
• Conforms to the Directive 97/23/EC (Pressure Equipment) This equipment is considered to be a pressure accessory and NOT a safety accessory as define
3 Working Principle The flow meter consists of a calibrated orifice and a conical float. The rising flow pushes the float to an equilibrium point. The
4 HORIZONTAL FLOW Figure 1 Figure 2 OPERATION Once the meter is installed, the regulator valve should be opened slowly and the fluid flow will move th
5 GAS MEASUREMENT. (Figure 3) THE WORKING PRESSURE IS OF MAXIMUM IMPORTANCE for correct GAS measurement readings as it directly affects the instrumen
Metering Tube 1.JAMMED FLOAT (M21/R, M21/N, M21/1, M21/3, M21/5, M21/7) A special tool is necessary to disassemble the flow meter float (1). This tool
7 12356Introduce the float into the top of the metering tube (the thin end should go first). Do this with care so as not to knock and damage the meter