Tecfluid CU Series Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messgeräte Tecfluid CU Series herunter. Tecfluid CU Series User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Instruction manual

Ultrasonic flowmeter CU100 Instruction manual R-MI-CU100 Rev.: 1 English version

Seite 2

10 3 CONVERTER INTERFACE The CU100 converter has a graphic LCD and a keyboard with 12 push buttons. The keyboard has ten numeric keys to introduce th

Seite 3

11 To change the access password of the programming menu, select "Programming" in the main menu and then "Password". 4 MAIN MENU

Seite 4

12 4 MAIN MENU When the "Password" option is selected, a screen to enter the new password appears. Once entered, the new password is asked

Seite 5 - 1 INTRODUCTION

13 5.1 Straight pipe run requirement In order to obtain the maximum reading stability and accuracy, the point where the transducers will be installe

Seite 6

14 5 TRANSDUCERS INSTALLATION 90º double elbow in one plane Valve Pump Disturbing element before the transducers 90º elbow 90º double elbow in diffe

Seite 7

15 90º double elbow in one plane Valve Disturbing element after the transducers Minimum distance between the second transducer and the ele-90º Elbo

Seite 8 - - mA (negative)

16 Proceed in the same way for the second strap. 5.3 Fastening of the transducers In order to fix the transducers it is necessary to calculate de op

Seite 9 - - Shield (negative)

17 5.6 Transducer You can choose the transducer model used in the installation. In order to access to the transducers installation menu, a password m

Seite 10 - 3 CONVERTER INTERFACE

18 5.7 Units User can choose between metric and imperial units. First, the outer diameter of the pipe must be entered. 5.8.1 Outer diameter The ou

Seite 11 - 4 MAIN MENU

19 5.8.4 Lining material If the pipe has an internal lining, in this screen the lining material can be selected. In the same way as in paragraph 5.

Seite 12

2 Thank you for choosing the flowmeter CU100 of Tecfluid S.A. This instruction manual allows the installation and operation of the ultrasonic flow m

Seite 13 - 5 TRANSDUCERS INSTALLATION

20 5.8.5 Lining thickness If the pipe has an internal lining, its thickness should be entered, in the same way as in the pipe thickness. 5.9 Liquid

Seite 14

21 For pipe diameters bigger than 350 mm, Z mode is recommended, which consists of placing the transducers on opposite sides of the pipe. The signal

Seite 15

22 5.11 Installation and distance between the transducers This screen shows the theoretical distance (Dt) at which the transducers should be placed

Seite 16

23 After placing the transducer, move the fastening "U" to it until the knurled screw matches the transducer top notch. If the alignment gu

Seite 17

24 5.13 Signal strength This screen allows to perform a fine adjustment of the position of the second transducer. To do so, the second transducer mus

Seite 18

25 6 CONVERTER PROGRAMMING By programming the converter the visualization and the outputs of the instrument can be configured. Turn on the converter

Seite 19

26 6.2 Units In this screen the units for the liquid velocity, the flow rate and the totalizer can be chosen independently. Note: Unit gal refers to

Seite 20

27 6.4 Flow rate This screen displays a submenu to modify different aspects related to flow reading, such as cut off, flow reversal, filtering and of

Seite 21

28 Only the flow rate indication of the display and the analog output are affected by the filter. The relay output and the totalizer act according to

Seite 22

29 6 When the read flow rate is stable, press the key OK and the instrument will save the value. The following step is to swap the cables of the tran

Seite 23

3 TABLE OF CONTENTS 1 INTRODUCTION ... 5 2 INSTALLATION OF THE

Seite 24

30 6.5.1 Relay 1 and relay 2 When selecting one of these screens we have access to program the flow rate at which the relays will change its state a

Seite 25 - 6 CONVERTER PROGRAMMING

31 Before doing a current calibration, be sure that the ammeter used for that is showing the real measure. 6.6 Default screen Flow rate, totalizer

Seite 26

32 6 The first screen after selecting “Totalizer” is the following: In this screen you can choose the flow direction for which you want to configure

Seite 27

33 6.8 Backlight This screen allows to select how long it will takes the backlight to turn off from the moment you release a key. 7 SOFTWARE VERSION

Seite 28

34 10 MAINTENANCE No special maintenance is required . For cleaning, a humid cloth can be used, and if necessary with a little neutral soap. Solven

Seite 29

35 11 ASSOCIATED SOFTWARE WINSMETER CU Most of the steps in the preceding paragraphs can be done by means of the device associated software Winsmete

Seite 30

36 If you have Internet access, answer “Yes, this time only”. If you do not have Internet access, answer “No, not this time”. If you answered “yes”,

Seite 31

37 11.2 Port connection In the "Port" section, choose the appropriate port for the converter. This will appear with the name of the port fo

Seite 32

38 Once the password is written, press "Enter" or the corresponding "Open", and the installation or programming tab will open as

Seite 33 - 8 SERIAL NUMBER

39 Changing the parameters of this screen, you can program the different functions of the equipment. As in the previous paragraph, to pass data to th

Seite 34 - 9 DEFAULT SCREEN

4 6 CONVERTER PROGRAMMING ... 25 6.1 Language ...

Seite 35

40 It is an intuitive tool to verify that the instrument has been installed and programmed correctly. By means of the "Record" button, you

Seite 36

41 12 TECHNICAL CHARACTERISTICS Accuracy ± 1,5% of reading ± 0,02 m/s Repeatability ± 0,25% of reading ± 0,01 m/s Velocity range 0,2 ...1

Seite 37

42 13 DIMENSIONS 13 DIMENSIONS

Seite 38

43 13 DIMENSIONS

Seite 39

44 14 TROUBLESHOOTING 14 TROUBLESHOOTING Problem Probable cause Solution The display shows an error informative message instead of the default scr

Seite 40

45 Annex A. Sound propagation velocity in some liquids (in m/s). Water 0 ºC 1403 10 ºC 1447 20 ºC 1482 30 ºC 1507 40 ºC 1526 50 ºC 1541 6

Seite 41 - TECHNICAL CHARACTERISTICS

46 Annex B. Sound propagation velocity in some solids (in m/s). B ANNEX B Pipe material Aluminium 3080 Brass 2150 Bronze 2270 Cast iron 2550 C

Seite 43

48 WARRANTY Tecfluid S.A. guarantees all the products for a period of 24 months from their sale, against all faulty materials, manufacturing or perfo

Seite 44 - TROUBLESHOOTING

5 1 INTRODUCTION The CU100 flowmeter is an electronic device that is based on ultrasonic signal transmission through a pipe where a liquid is flowing

Seite 45

6 2.1 Electrical connection For the electrical connection, the CU100 converter has four terminal strips. To help in the wiring of the equipment, the

Seite 46

7 2.1.1 Power supply wiring Before starting the installation of the instrument, check that the supply voltage available is the same as marked on the

Seite 47 - ANNEX C

8 2 The relay outputs are SPDT relays with potential free contacts (see characteristics in page 40). The state of the relay contacts corresponds to t

Seite 48

9 The mA output is galvanically isolated. It can be active (which means that the receiver device connected to it should have a passive input) or pass

Kommentare zu diesen Handbüchern

Keine Kommentare