Instructions Manual R-MI-XT5 Rev.: 2 English version XT5 /XT5H Converter for Flomid/Flomat Sensors instrumentation for fluids
10 In the case of using a par-pos cable, which has two shielded pairs of cables (one for the electrodes and one for the coils) there is only one shiel
11 3.1.3 Nominal diameter In this screen the nominal diameter (internal diameter) in millimetres of the sensor, or the pipe size in the case of a Flom
In the first screen the flow rate for the 4 mA point (lower range) is configured. In the next screen the flow rate for the 20 mA point (upper range) i
3.2.3 Pulse output In this screen, there are two options. a) Frequency output (Hz). This is meant for transmitting the instantaneous flow rate to a
14 Only the flow rate indication of the display and the analog output are affected by the filter. The pulse output and the totalizer act according to
In an XT5H converter, if during a configuration sequence a HARTTM command, which should be attended, is received, the local configuration sequence wi
In order to see the software version and return to the normal working screen, touch any key. 3.5 Empty pipe indication When the XT5 detects that the
Summary of the principal communication characteristics: 17 Manufacturer, Model and Revision Tecfluid S.A., XT5 converter, Rev. 1 Device type Transm
7 MAINTENANCE No special maintenance is required . For cleaning, a humid cloth can be used, and if necessary with a little soap. Solvents or other ag
9 TECHNICAL CHARACTERISTICS 9.1 Power supply 230, 240, 115, 24 VAC 50, 60 Hz . 24 VDC. Power consumption: ≤ 5 VA 9.2 Analog output 4-20 mA. Active
INDICE 1. INTRODUCTION ... 4 2. ELECTRICAL CONNECTON ...
10 DIMENSIONS COMPACT MODEL WALL MOUNTING MODEL 20
11 TROUBLESHOOTING Problem Probable cause Remedy The flow rate is not displayed. In its place there are dashes The pipe is empty. Make sure that th
12 PROGRAMMING DIAGRAM Problem Probable cause Remedy The display is blank Fused fuse. Change the fuse. The analog output gives always 4 mA or 20 m
dP: Number of decimal digits LO: Equivalent flow rate to 4 mA in the current loop UP: Equivalent flow rate to 20 mA in the current loop P/U: Pulse
WARRANTY Tecfluid S.A. GUARANTEES ALL ITS PRODUCTS FOR A PERIOD OF 24 MONTHS, after consignment, against all defects in materials and workmanship. Thi
3 7. MAINTENANCE …... 18 7.1. Fuse ……………………………..……………………………...
1 INTRODUCTION The XT5 converter is an electronic equipment that adapts to the Flomid and Flomat series electromagnetic flow sensors. The microprocess
2.1 Power supply connection Terminal AC Power supply DC Power supply Earth Earth N Neutral 0 V (-) L Phase 24 V (+) 5 It is import
6 To configure the analog output type (active or passive) there is a slide switch situated just behind the terminal strip. For the passive mode the sw
7 mA: Ammeter
2.3 Pulse output connection Terminal E Emitter. C Collector. 8 2.4 Connection examples The two usual ways to connect the alarm outputs are N
9 PNP connection 2.5 Sensor connection When the electronics is mounted separate from the sensor, the interconnection is made through a cable gland. Th